What is the usual method of monitoring the remaining contents in a gas-filled cylinder?


A) Weigh the cylinder.
B) Read the pressure gauge.
C) Compute the gas density.
D) Read the cylinder label.

E) A) and D)
F) A) and C)

Correct Answer

verifed

verified

The gauge on an E cylinder of O2 reads 800 psig. About how long would the contents of this cylinder last, until completely empty, at a flow of 3 L/min?


A) 1 hour 15 minutes
B) 1 hour 45 minutes
C) 2 hours 10 minutes
D) 2 hours 40 minutes

E) A) and B)
F) All of the above

Correct Answer

verifed

verified

What do the components of a liquid O2 bulk storage tank include? I.inner and outer steel shells, separated by a vacuum II.safety system that vents O2 if warming occurs III.vaporizer system that converts liquid O2 to gas IV.pressure-reducing valves to lower pressure to 50 psig


A) I, II, and III
B) II and IV
C) I, II, III, and IV
D) III and IV

E) All of the above
F) C) and D)

Correct Answer

verifed

verified

When minimal fluctuations in pressure and flow are critical factors, which of the following pressure-reducing valves would be the best choice?


A) single-stage preset pressure-reducing valve
B) multiple-stage pressure-reducing valve
C) uncompensated Thorpe tube flowmeter
D) Bourdon adjustable pressure regulator

E) A) and C)
F) C) and D)

Correct Answer

verifed

verified

The pressure of O2 or air in a bulk supply system is reduced to what standard working pressure?


A) 10 psig
B) 14 psig
C) 25 psig
D) 50 psig

E) B) and C)
F) A) and D)

Correct Answer

verifed

verified

Compared to portable cylinders, a centrally located, piped-in bulk O2 system has which of the following advantages? I.no need for a backup or reserve system II.more convenient and less hazardous than cylinders III.oxygen already reduced to its working pressure IV.less risk of the O2 supply being depleted


A) I, II, and III
B) II and IV
C) III and IV
D) II, III, and IV

E) B) and C)
F) C) and D)

Correct Answer

verifed

verified

Compressed gas cylinders are manufactured from which of the following?


A) copper or brass
B) seamless steel
C) seamless aluminum
D) fiberglass

E) All of the above
F) C) and D)

Correct Answer

verifed

verified

In clinical practice, how is a positive identification made of the contents of a medical gas cylinder?


A) noting the color of the cylinder
B) inspecting the cylinder threads
C) reading the cylinder label
D) collecting and analyzing a gas sample

E) A) and B)
F) A) and C)

Correct Answer

verifed

verified

During inspection of the shoulder of a compressed gas cylinder, you note a plus sign (+) next to the test date. This indicates what about the cylinder?


A) It is made of spun aluminum, not steel.
B) It only requires a 10-year DOT inspection.
C) It can be filled to 10% above its service pressure.
D) It has a high coefficient of elastic expansion.

E) B) and C)
F) A) and B)

Correct Answer

verifed

verified

What is the primary purpose of indexed connector systems?


A) to prevent inadvertent misconnections between equipment
B) to speed making connections between equipment
C) to allow U.S. equipment to "mate" with foreign equipment
D) to provide universal connections among all equipment

E) All of the above
F) A) and B)

Correct Answer

verifed

verified

The needle valve of a flowmeter with a ball and float tube is completely closed. When you plug the flowmeter into a gas source, the float does not move. Which of the following can be concluded?


A) The device is not calibrated at 50 psi.
B) The device is a Bourdon-gauge flowmeter.
C) The device is pressure compensated.
D) The device is affected by gravity.

E) A) and B)
F) B) and C)

Correct Answer

verifed

verified

If you have to deliver O2 to a patient directly from a bedside outlet station, which of the following devices would you select?


A) Thorpe tube flowmeter
B) Bourdon type gauge
C) pressure-reducing valve
D) medical gas regulator

E) B) and D)
F) B) and C)

Correct Answer

verifed

verified

What is the indexed safety system for threaded high-pressure connections between large compressed gas cylinders and their attachments?


A) Pin-Indexed Safety System (PISS)
B) Diameter-Index Safety System (DISS)
C) American Standard Safety System (ASSS)
D) CGA System

E) A) and B)
F) None of the above

Correct Answer

verifed

verified

Unlike CO2 or N2O, O2 cannot be stored as a liquid at room temperature but must be kept in special cryogenic containers. Why is this so?


A) O2 is 20 times less diffusible than CO2, N2O, or cyclopropane.
B) O2 has a critical temperature well below room temperature.
C) O2 has an extremely high gas solubility coefficient.
D) O2 has a boiling point below well below room temperature.

E) A) and B)
F) A) and C)

Correct Answer

verifed

verified

The gauge on a G cylinder of O2 reads 400 psig. About how long would the contents of this cylinder last, until completely empty, at a flow of 5 L/min?


A) 2 hours 40 minutes
B) 3 hours 10 minutes
C) 4 hours 30 minutes
D) 9 hours 15 minutes

E) None of the above
F) A) and D)

Correct Answer

verifed

verified

What is the PISS index hole position for O2?


A) 1 and 5
B) 2 and 5
C) 1 and 6
D) 4 and 6

E) A) and C)
F) A) and D)

Correct Answer

verifed

verified

The measured pressure in a liquid-filled cylinder is equivalent to which of the following?


A) the force necessary to compress its volume within the cylinder
B) its filling density divided by the cylinder gas factor
C) the pressure of the surface vapor at any given temperature
D) the weight of the liquid divided by the density of water

E) A) and B)
F) B) and D)

Correct Answer

verifed

verified

Which of the following is false about ASSS standards?


A) They provide specifications for more than 60 gases and mixtures.
B) They apply only to cylinders sizes F through H/K.
C) They apply to high-pressure connections.
D) They provide a separate connector for all gases.

E) A) and B)
F) A) and C)

Correct Answer

verifed

verified

What is the key difference between small compressed gas cylinders (sizes AA to E) and their larger counterparts (sizes F to K) ?


A) Small gas cylinders do not undergo regular DOT testing.
B) Small gas cylinders are always filled to lower pressures.
C) Small gas cylinders cannot be used for anesthetic gases.
D) Small gas cylinders use a yoke (not threaded) connector.

E) C) and D)
F) B) and C)

Correct Answer

verifed

verified

The NFPA standard for bulk liquid O2 systems requires that the reserve supply be which of the following?


A) fixed cylinder bank of 75 H cylinders
B) at least one large liquid O2 cylinder
C) equal to the average gas use of 1 day
D) at least 3000 cu/ft of gaseous O2

E) None of the above
F) A) and D)

Correct Answer

verifed

verified

Showing 21 - 40 of 80